Siemens Siremobil Compact L C-Arm Specifications
|
|
C-arm: |
- Orbital movement 130° (-40° to + 90°)
- Angulation ± 190°
- Horizontal movement 20 cm
- Immersion depth 73 cm
- Swivel range ± 12.5°
- Vertical travel 45 cm (17.7″) motorized
- Source – I.I. distance 100 cm (39.4″)
- Free space 78 cm (30.7″)
|
|
X-Ray Generator/Tube: |
- Converter control frequency 15 kHz to 30 kHz
- KV range 40 kV to 110 kV
- Fluoroscopy 0.2 mA to 8.9 mA
- Digital radiography 0.2 mA to 12.2 mA
- Cassette exposures Max. 20mA
|
|
Single Tank with Stationary Anode: |
- Focal spot nominal value 0.6
- Nominal voltage 110 kV
- Optical anode angle 9°
- Inherent filtration ≥ 3 mm AI equivalent
|
|
Collimator System: |
- Iris collimator for concentric collimation
- Semi-transparent slot diaphragm for symmetric, radiation-free collimation, with unlimited rotation
|
|
TV System: |
- VIDEOMED DC with CCD sensor
- TV standard 625 lines at 50 Hz, 525 lines at 60 Hz
- Digital image rotation ± 220°
|
|
X-Ray Image Intensifier: |
- Metal-enamel technology (23cm 9″ l.l.), Metal-ceramic technology (17 cm, 7″ l.l.) with Mu-metal shield
- Precision electron optics with minimal image distortion and consistent high resolution across the entire image field
- Cesium-iodide input screen for minimum quantum noise and excellent resolution
- Anti-glare output screen with scattered light trap for high contrast dynamics and prevention of scattered light effects
- High-transparency input window
- SIRECON 23-2 HDR-C:
- Nominal diameter 23 cm (9″)
- Zoom format 15 cm (6″)
- Grid PB 8/40, f 90
- SIRECON 17-2 HDR-C
- Nominal diameter 17 cm (7″)
- Zoom format 10 cm (4″)
- Grid PB r17 N708/40, f 85
|
|
Operating Modes: |
- Fluoroscopy with Last Image Hold
- Pulsed Fluoroscopy with Last Image Hold
- Digital radiography
- Digital Subtraction Angiography
- Roadmap
- Selection from 5 application-specific fluoroscopy and radiography curves for the individual operating modes
|
|
Image Processing: |
- Recursive filtration, summation or motion-dependent noise reduction
- Spatial frequency filtration for edge-enhanced image display, contrast enhancement
- Digital image rotation with subsequent processing
- Positive/negative image version
- Left/right and top/bottom image reversal
- Zoom, Roam
- Subtraction, Roadmap, Pixelshift, Remask, Landmark
- Up to 6 lookup tables for contrast enhancement
- Three-stage edge enhancement
- Electronic shutters, horizontal and vertical
|
|
|
Trolley Options
|
|
|
- Suitable for one 15″ b/w or two 15″, 17″, 18″ b/w displays
|
|
15″ TFT b/w Display: |
- Screen size 15″ (38 cm)
- Image matrix 1024 x 768
- Maximum brightness 250 cd/m2
|
|
17″ TFT b/w Display: |
- Screen size 17″ (43 cm)
- Image matrix 1280 x 1024
- Maximum brightness 250 cd/m2
|
|
18″ TFT b/w Display: |
- Screen size 18″ (46 cm)
- Image matrix 1280 x 1024
- Maximum brightness 600 cd/m2
|
|
Options: |
- Multiformat camera
- Printer
- Laser camera connection
- System-integrated laser targeting device on tube side
- Detachable laser aimer on image intensifier side
- Sterile cover for C-arm, X-ray tube and image intensifier
- Integrated dose measurement chamber
- Memory extension
- DICOM
- DSA
- CD writer
|
|
Power Requirements: |
- 100V,110V,120V,127V,200V,230V,240V; (±10%), 60/60 Hz (±1 Hz)
|
|
Environmental Conditions: |
- Temperature range: +15°C to + 35°C
- Relative humidity: 15% to 75%, non-condensing
- Barometric pressure: 700hPa to 1060 hPa
|
|
Weight: |
- C-arm chassis 257 kg
- Display trolley with one display 112 kg
- Display trolley with two displays 138 kg
|