GE OEC 9900 Elite Mobile C-Arm Specifications
|
|
Generator: |
- 60kHz high frequency
- 15kW full-wave
- Up to 120kVp
- Up to 75mA for radiographic film exposure
- Continuous high level fluoro (HLF) up to 20mA
- Digital spot up to 75mA
|
|
X-Ray Tube: |
- Rotating anode X-ray tube
- 0.3mm and 0.6mm focal spots
- Anode heat capacity: 300,000 HU (per IEC 60613)
- Anode cooling rate: 85,000 HU/min
- Housing heat capacity: 1,600,000 HU
- Standard housing cooling: 15,000 HU/min
- Passive housing cooling system: 22,500 HU/min (Standard on 12 inch (31 cm) I.I. systems. Optional on 9 inch (23 cm) systems
|
|
Digital Image Rotation |
- Digitally adjusts image display
- Automatic image update
- Image rotation
- Image reversal (side-to-side)
- Image invert (top-to-bottom)
- Image positioning without additional exposure
|
|
PreViewTM Collimator: |
- On-screen collimator position indication
- PreViewTM iris collimator
- PreViewTM Tungsten rotatable double leaf collimator
- Adjusts collimators without X-ray exposure
|
|
Fluoro Mode: |
- kVp range: 40 – 120
- mA range: 0.2 – 10 mA normal mode 1.0 – 20 HLF (high level fluoro)
- Auto and manual flouro modes
- AutoTrakTM ABS varies mA, kVp and camera gain
|
|
Pulsed Fluoro Mode: |
- kVp range: 40 – 120 kVp
- mA range: 0.2 – 10
- Pulse rate: 1, 2, 4, 8
- Pulse width: 25 or 50 milliseconds
- AutoTrakTM ABS, mA, kVp, camera gain
- Reduces X-ray to patient and operator
|
|
High Level Pulsed Fluoro: |
- kVp range: 40 – 120 kVp
- mA range: 0.2 – 40
- Pulse rate: 1, 2, 4, 8
- Pulse width: 25 or 50 milliseconds
- AutoTrakTM ABS, mA, kVp, and camera gain
|
|
Digital Spot Mode: |
- kVp range: 40 – 120
- mA range: Up to 75
- Automatic exposure termination and automatic image save Radiographic Mode
|
|
Radiographic Mode |
- mA range: up to 75
- mAs range: up to 300
- Computer controlled exposure time
- 10″ x 12″ (24cm x 30cm) for 9″ I.I.
- 14″ x 14″ (35cm x 35cm) for 12″ I.I.
|
|
|
Video Imaging System
|
|
9″ Image Intensifier (optional): |
- Tri-mode 9″/6″/4.5″ (23cm/15cm/11cm) image intensifier
- Minimum central resolution (at the monitor):
- 9″ (23cm): 2.2 lp/mm
- 6″ (15cm): 3.0 lp/mm
- 4.5″ (11cm): 3.5 lp/mm
- DQE: 65% (typical)
|
|
12″ Image Intensifier (optional): |
- Tri-mode 12″/9″/6″ (31cm/23cm/15cm) image intensifier
- Minimum central resolution (at the monitor):
- 12″ (31cm): 1.6 lp/mm
- 9″ (23cm): 2.2 lp/mm
- 6″ (15cm): 2.6 lp/mm
- DQE: 65% (typical)
|
|
Precision imaging with Dynamic Range Management (DRM) enhances features of interest while attenuating background noise: |
- Preset Imaging Profiles
- 9800
- General
- Orthopedic
- Spine
- Cardiac
- Vascular
- Bolus Chase
|
|
AutoTrakTM Automatic Brightness Stabilization (ABS): |
- Automatically seeks the subject anatomy anywhere within the imaging field and selects the optimum imaging technique
- Automatically adjusts to anatomical size and location
- Provides uniform image quality throughout entire image
- Simplifies operation
|
|
Image I.Q.: |
- Smart Window
- Dynamically senses the collimator position and automatically adjusts brightness and contrast to produce high image quality
- Smart Metal
- Allows user to adjust automatic brightness and contrast sensitivity levels to metal
- Provides optimum image quality even when metal is introduced to the field
- Tungsten Collimator
- Denser collimator limits X-ray exposure area
- Reduces scatter radiation
- Improves image detail
|
|
Video Camera: |
- High resolution 1k x 1k CCD camera
- Full frame capture
- Motorized rotation
- On-screen orientation indicator (realtime feedback without fluoro)
- Left-right image reversal
- Top-bottom image invert
|
|
Video Display: |
- Dual 18″ (46cm) display anti-glare, LCD flat panel monitors mounted on an articulating arm
- 22″ horizontal travel
- 7° up/10° down
- Monitors viewable from all four sides of workstation
- Horizontal and vertical viewing angle 170°
- 700 CD/M2 maximum brightness
- 1280 x 1024 high resolution monitors
- Ambient room-light compensation
- Integrated color monitor for display of VGA, DVI, DVI-D, S-VHS, and SDI-SD formats (optional)
|
|
|
Configurations
|
|
GSP Platform: |
- 1k x 1k x 16 bit image processing
- Preset Imaging Profiles
- 9800
- General
- Orthopedic
- Spine
- Noise filter with on-screen indicator
- Minimal difference spatio-temporal noise filter (MDST)
- Real time dynamic range management (DRM)
- Automatic digital brightness and contrast control
- Manual digital brightness and contrast control
- Negate mode
- Save and auto-save feature
- Swap and auto-swap feature
- Patient information
- Examination list
- Customized patient information
- Customize functions
- Workstation set-up
- Mainframe set-up
- Patient information set-up
- Date/time set-up
- DICOM interface set-up
- Last image hold
- 63 image storage
- CD/DVD burner with DICOM viewer for displaying images on PC platforms
- Integrated DICOM interface (store, print, worklist, and query/retrieve)
- HIPAA SecureView®
- Password protection
- Blank screen function
- Delete all
|
|
ESP Platform: |
Includes all the GSP features plus:
- 1000 image storage
- Zoom and roam function
- Image annotation
- Measurement software
|
|
ESP Platform with 15fps Digital Disk: |
Includes all the ESP features plus:
- 15fps Dynamic digital disk (60 Hz) 12fps (50 Hz)
- Recording/playback rate: 1, 2, 4, 8, 15fps (60 Hz) 12fps (50 Hz)
- Automatic image playback capability
- Frame-by-frame review
- Recording time: 10 minutes @ 15fps (60 Hz) 12 fps (50 Hz)
|
|
PMCare Platform: |
Includes all the ESP features plus:
- Real-time subtraction
- Digital subtraction (DSA) angio
- Peak opacification
- 8fps Dynamic digital disk
- Recording/playback rate: 1, 2, 4, 8fps
- Recording time: 5 minutes @ 8fps
- Automatic image playback
- Frame-by-frame review
|
|
Basic Vascular Platform: |
Includes all the PMCare features plus:
- Vascular preset imaging profile
- Roadmapping
- Re-registration
- Variable landmarking
- Mask save/recall
|
|
Vascular MTS Platform: |
Includes all the Basic Vascular features plus Motion Tolerant Subtraction (MTS) plus:
- Bolus chase preset imaging profile
- Digital cine pulse mode
- 30 pulses/sec 60 Hz (25 pulses/sec 50 Hz)
- Up to 150mA
- 10ms pulse width
- 30fps Dynamic digital disk 60 Hz (25 pulses/sec, 50 Hz)
- Recording/playback rate: 1, 2, 4, 8, 15, 30fps 60 Hz (1, 2, 4, 8, 12, 25, 50 Hz)
- Recording time: 10 minutes @ 30fps 60 Hz (time depends on record frame rate) (25 fps, 50 Hz)
|
|
Cardiac Platform with Interventional Vascular Capability: |
Includes all the Vascular MTS features plus:
- Cardiac preset imaging profile
- Super “C” configuration (9″/23cm I.I. only)
- Single leaf curved collimator
- Three pedal footswitch
|
|
|
Additional Features
|
|
Hardcopy Options: |
- Integrated film/paper printer
- No film developing required
- 8″ x 10″ (20.3 cm x 25.5 cm) laser quality film/paper
- Multi-format, 1, 2, 4 on 1
- Multi-copy capability
- Thermal printers
- Integrated DICOM interface (storage class/print class/query work-list)*
|
|
User Interface: |
- Entire system is computer controlled and software upgradeable
- Touchscreen control simplifies operation
- Automated system operation requires minimum operator interface
- Multi-functional controls
- Footswitch
- IR remote (optional)
- Hand-held control
- Simplified keyboard with integrated touchpad
- Multi-purpose image directory
- Retrieve and review images
- Compose hardcopy films
- Copy images
- X-ray dose summary
|
|
Uninterruptible Power: |
|
|
Electrical Specifications: |
- Input power (60 Hz or 50 Hz) 9900 Elite
- 120V @ 12 A
- 200V, 220V, 230V, 240V @ 10 A
|